1 March 2009 · research
Rapid Nondestructive Measurement of Root Dentin Moisture
Qiang Zhu, Jin Jiang, Kamran E. Safavi, Larz S.W. Spångberg, Takashi Komabayashi
Published in: Author manuscript, OpenCommons@UConn (2009)

Experimental study: a digital moisture meter for assessing dentin without destroying the specimen (OpenCommons@UConn, 2009)
Abstract
Objectives
To evaluate whether root dentin moisture can be measured rapidly and nondestructively with a digital grain moisture meter, given that moisture matters for adhesive cementation and strength and no rapid nondestructive method existed.
Materials and methods
Extracted single-rooted mandibular teeth were sectioned at the cemento-enamel junction. Root dentin moisture was measured on six grain settings, with five repeats each. As the reference, the change in specimen weight before and after drying was compared with the meter readings.
Results
Each nondestructive measurement took less than 30 seconds. Storage for 24 hours at 37 °C and 100% humidity did not restore the original dentin moisture. Five of the grain settings showed high validity.
Conclusions
A digital grain moisture meter is suitable for rapid, nondestructive assessment of root dentin moisture.
Dentin moisture affects both bonding and tooth strength, yet there has been no rapid way to assess it without destroying the specimen. This study tests whether a digital grain moisture meter can be used to measure root dentin moisture in vitro.
Keywords
Dentin moisture, Nondestructive method, Impedance, Dentin bonding, Dentin strength
Related articles

6 August 2020 · Review · Diagnosis & clinic
Pathogenesis, diagnosis and management of dentin hypersensitivity: an evidence-based overview for dental practitionersReview: risk factors, the hydrodynamic theory and strategies for the differential diagnosis and management of dentin hypersensitivity (BMC Oral Health, 2020)